Kullanım alanları: Hidrokarbon prosesleri için inert örtülü gazların izlenmesi, ilaç veya kimya sanayi için inert gaz; biyogaz, atık çöp ve sindirme tesisleri, demir-çelik sektöründe fırın gazı kontrolü, katalizör yenilenmesi, azot üretimi, oksijen üretimi.
Ürün Özellikleri
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Teknik Özellikler
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Performans |
|
| Measurement Technology | Thermo-paramagnetic oxygen sensor |
| Gas | Process and non-condensing sample with particles <3µm |
| Measurement Range |
Selectable from 0-1% up to 0-50% O2 and 80/90 - 100% |
| Display Resolution | 0.01% |
| Display Type | Backlit LCD |
| Accuracy Excluding Suppressed Zero Ranges |
< ±1% of span or ±0.02% O2 whichever is greater |
| Accuracy For Suppressed Zero Ranges (80/90 - 100%) | < 0.2% O2 |
| Response Time (T90) with High Speed Response (HSR) Enabled | < 15 seconds |
| Repeatability |
±0.2% of span or 0.01% O2 whichever is greater |
| Linearity | ±0.5% of span or 0.05% O2 whichever is greater |
| Zero Stability | ±0.25% of span per month |
| Span Stability | ±0.25% of span per month |
| Sample Flow Rate | 100 to 600 ml/min |
| Sample Flow Effect (calibrated at 300 ml/min) |
< 1% of span for flows: 100 to 600 ml/min |
| Sample Pressure |
0 to 3 barg inlet with atmospheric vent |
| Sample Temperature | 0 to +55°C max (+45°C for standard cell) |
| Sample Cell Temperature | Standard +50°C (Optional +55°C and +60°C) |
| Background Gas | Unit can be calibrated in a user-defined background gas to increase the accuracy |
|
|
| Analog Inputs | 2 off 4-20 mA inputs One for an external sensor that can be displayed on the screen One to act as an active compensation for the process conditions |
| Analog Outputs | 2 off 4-20 mA outputs (isolated) |
| Output Ranges |
Primary range is set to the calibrated range of the instrument The second is user selectable within the primary range |
| Alarms | 2 off single pole changeover (SPCO) relays for O2 concentration (250 V, 5 A max) |
| Datalogging | The user can use the digital communications to log the output from the analyzer The unit will store 40 alarm points and the min/max O2 concentrations with date and time stamp |
| Digital Communications | MODBUS RTU over RS485 Protocol |
| Power Supply |
24 V DC; 1.5 A max |
|
Çalışma Koşulları |
|
| Ambient temperature | Dependent on o-ring selection: Silicon: –40°C to +55°C Viton®: –15°C to +55°C Ekraz: –10°C to +55°C |
|
Mekanik Özellikler |
|
| Warm Up Time | < 30 minutes |
| Stabilization Time | 15 minutes |
| Dimensions |
234 x 234 x 172mm (w x d x h) |
| Weight | 9.7kg |
| Wetted Materials | 316 & 430F stainless steel, boroscillicate glass, platinum, 3M 2216, plus O-ring |
| O-Ring Materials |
Viton, Silicone or Ekraz |
| Gas Connection | 1/4” NPT, 1/4” tube or 6mm tube |
| Ingress Protection | IP66, NEMA 4X |
|
|
| ATEX | II 2GD Ex d IIB +H2 T3 Gb Ex tb IIIC T137°C Db IP66 |
| IECEx | Ex d IIB +H2 T3 Gb Ex tb IIIC T137°C Db IP66 |
| Temperature Ranges for ATEX and IECEx as per O-ring Type |
Silicon: Ta = -40°C to +55°C Viton®: Ta = -15°C to +55°C Ekraz: Ta = -10°C to +55°C |
| CCSAUS | Class I, Division 1, Groups B,C,D |
| INMETRO | Ex d IIB + H2 T3 Gb Ex tb IIIC T137°C Db IP66 |
| TC TR EX |
1 Ex d IIB + H2 T3 Gb |