Ana Ünite Ölçme Parametreleri
 | 
| Channel Configuration | 
Single and dual channel | 
| Moisture Content | 
ppmw | 
| Temperature | 
°C and °F | 
| Water Dew Point | 
°C and °F
  | 
Sample Flow Rate
  | 
Status indication: Low-Good-High (<0.1 to >0.3 l/min) | 
 
Sensör Teknolojisi
 | 
| Moisture Sensor Technology | 
Michell Ceramic Moisture Sensor | 
| Calibration Range | 
-100 to +20°Cdp | 
| Measurement Range | 
-120 to +30°Cdp (0.001 ppmw to saturation) | 
| Accuracy | 
Dew point: ±1°C between -59.9 & +20°Cdp Moisture content: ±10% of reading Dew point: ±2°C between -60 & -100°Cdp Moisture content: ±20% of reading Analysis temperature: Accuracy ±0.2°C | 
| Resolution | 
0.1°C between +20 and -80°Cdp 1°C between -80 and -100°Cdp
  | 
Temperature coefficient
  | 
Algorithm compensation | 
| Operating Pressure | 
Max. 50 barg
  | 
Sample Flow
  | 
Up to 1 l/min (0.1 - 0.2 l/min recommended) | 
 
HMI
 | 
| Keypad/Interface | 
Capacitive touch-screen through glass | 
| Display | 
Vacuum fluorescent | 
| Datalogging | 
A rolling maximum of 150 data logs is available Each log records time, date, moisture and temperature values for each channel 5 minutes minimum and 60 minutes maximum logging intervals can be set by the user
  | 
| Communications and Output | 
Two non-isolated 4-20 mA per measurement channel 500 Ω maximum load Range and parameter settable by user Modbus RTU @ 9600 baud-rate  Alarms: two volt free contacts per channel; one process value and one instrument status Also available via Modbus communications
  | 
 
Koruma
 | 
| Type | 
Flameproof EExd | 
| Construction | 
Cast copper-free aluminium (IP66/NEMA 4) | 
| Finish | 
Chromate primer, Polyester P9010 powder coated (black). Meets BS3900 | 
 
Genel
 | 
| Sample Connections | 
1/4" NPT (female) | 
| Weight | 
22.5kg | 
| Operating Supply Voltage | 
90 to 260 V AC, 50/60 Hz, 180 W Main Unit | 
| Operating Environment | 
-20 to +60°C up to 95% RH Indoor or Outdoor, shaded location | 
 
Sertifikasyon
 | 
| Hazardous Area Certification | 
ATEX Certification: II 2 G Code: EEx d IIB + H2 T4 (135°C) IP66 Tamb -40 to +60°C CCSAUS approved: Div 1, Class 1, Group B,C,D  GOST Ex | 
| Pattern Approval | 
GOST-R, GOST-K | 
 
Kapalı Güvenli Alan Montajı için Opsiyonel Uzaktan Arabirim
 | 
| Host Capacity | 
Up to 31 Michell EExd Process Area Installation Instrumentation Range analysers in any combination of models via Modbus | 
| Display | 
Backlit color 15cm LCD | 
| User Interface | 
Integral mouse, on-screen keyboard | 
| Function | 
View and modify analyser operating settings Record parameters to file and view virtual chart recorder analyser diagnostics | 
| Ports | 
LAN, USB2, Modbus Re-transmission | 
| Data Download | 
CSV (comma separated variable) format for easy import to ExcelTM | 
| Remote Access | 
OPC (OLE for Process Control) over a LAN Embedded web page/unique IP address for PC access Remote data reading via Modbus RTU | 
| Power Supply | 
100-280V, 50/60Hz | 
Plant Control Integration
  | 
ActiveX components to be used by software programmers for integration of Modbus protocol into a general plant software system |