 Kullanım alanı: İlaç sanayi, tarım, güç, enerji, nükleer, kağıt prosesi, ulaşım, denizcilik, metalurji, yiyecek ve içecek endüstrisi. 
 Ürün Özellikleri 
 | 
| 
 
 
 
 Ürünler  | 
![]() DT282/284-XP  | 
![]() DT292/294-XP Display  | 
| Performans Özellikleri | ||
| RH Measurement Range | 
0 to 100% RH | 
|
| RH Accuracy @ 23°C | ±1.25% RH (5 to 95% RH) | 
|
| RH Thermal Coefficient | 
<0.03% RH/°C typical | |
| RH Measurement Response Time | 
<10 seconds | |
| RH element | 
H8000 | |
| RH Long Term Stability | 
±1% RH per year | |
| Temperature Measurement Range | 
–40 to +85°C (DT282/292) –40 to +140°C (DT284/294)  | 
|
| Temperature Accuracy | 
±0.2°C | 
|
| Temperature Analog Output Accuracy | 
±0.2°C pro-rata per 100°C output signal span | |
| Temperature Resolution | 
0.01°C | |
| Temperature Technology | 
PT100 1/3 DIN | 
|
| Calibration | 
Traceable 5-point calibration certificate | 
|
| Elektriksel Özellikler | ||
| Output signals | 0–1 V, 0–5 V, 0–10 V, 4–20 mA, 0–20 mA | 
|
| Digital output signal | 
RS485 ModBus RTU 2-wire | 
|
| mA load resistance | 
Max 500 Ω over voltage supply range | 
|
| Supply voltage range | 
15 to 21 V AC or 18 to 30 V DC | 
|
| Power consumption | 1.7 W max | 
|
| Calculated 3rd output scales/range | 
Dew point: –40 to +100°C Mixing ratio: 0 to 500 g/kg Absolute humidity: 0 to 600 g/m3 Specific enthalpy: 0 to 1500 KJ/kg Frost point: –50 to +10°C  | 
|
| CE conformity | 
2004/108/EC, 94/9/E | 
|
| Communication kit (optional) | 
Digicor2 280/290 software, USB to RS485 interface cable (Digicor2-CK-01) | 
|
| Ekran Özellikleri | ||
| Display type | 
LCD, 2 lines x 16 characters | |
| Display configuration | Digicor2 communication kit and display keypad | 
|
| Display configuration variables | Contrast, temperature averaging, RH averaging, temperature units | 
|
| Display configuration output variables | 
Output RH signal, output temperature signal, calculated 3rd output, ModBus set-up | 
|
| Çalışma Koşulları | ||
| Operating temperature | 
Probe: Housing: Storage:  | 
–40 to +85°C (–40 to +185°F) (DT282/292) –40 to +140°C (–40 to +284°F) (DT284/294) –30 to +70°C (–22 to +158°F) –40 to +70°C (–40 to +158°F)  | 
| Mekanik Özellikler | ||
| Ingress protection | 
IP65 (NEMA 4 level) | 
|
| Material | Probe: Housing: Filter assembly (DT282/292): Filter assembly (DT284/294):  | 
Stainless steel (DT282/292) Stainless steel, PEEK (DT284/294) Aluminum die casting Stainless steel sintered filter Stainless steel sintered filter with Teflon Slotted stainless steel cap with stainless steel mesh filter Slotted stainless steel cap with PTFE filter  | 
| Dimensions | 
Probe: Housing:  | 
ø19mm, 250mm (9.8”) or 500mm (19.7”) length (DT282/292) ø12mm, 250mm (9.8”) or 500mm( 19.7”) length (DT284/294) 120 x 120 x 49.5mm (4.7 x 4.7 x 1.9”)  | 
| Weight | 
450g (0.99lbs) (approx) | 
|
| Electrical connections | Internal screw terminals | 
|