| Performans
 | 
| Measurement Range | -100 to +20°C dew point -110 to +20°C dew point
 | 
| Accuracy 
 | ±2°C dew point | 
| Response Time | 5 mins to T95 (dry to wet) | 
| Repeatability 
 | 0.5°C dew point 
 | 
| Calibration | Traceable 7-point calibration certificate | 
| 
 Elektriksel Özellikler | 
| Output Signal 
 | 4-20 mA (2-wire connection, current source) User-configurable over range
 | 
| Output | Dew point, moisture content in ppmV | 
| Analog Output Scaled Range 
 | Dew point: -100 to +20°C Moisture content in gas: 0 - 3000 ppmV
 Non-standard available upon request
 | 
| Supply Voltage | 12 to 28 V DC | 
| Load Resistance 
 | Max 250 Ω @ 12 V (500 Ω @ 24 V) | 
| Current Consumption 
 | 20 mA max | 
| CE Marked 
 | Certified | 
| Çalışma Özellikleri
 | 
| Operating Temperature | -40 to +70°C | 
| Operating Pressure | 45 MPa (450 barg) maximum | 
| Overpressure Rating 
 | x2 operating pressure 90 MPa (900 barg) 
 | 
| Flow Rate 
 | 1 to 5 Nl/min mounted in standard sampling block 0 to 10 m/sec direct insertion
 
 | 
| Temperature Coefficient | Temperature compensated across operating temperature range 
 | 
| 
 Mekanik Özellikler | 
| Ingress Protection Sensor 
 | IP66 in accordance with standard BS EN 60529:1992 NEMA 4 in protection accordance with standard NEMA 250-2003
 | 
| Hazardous Area Certificates 
 | ATEX - II 1 G Ex ia IIC T4 (-20°C ≤ Ta ≤ +70°C) FM - IS / I / 1 / ABCD / T4 Ta = +70°C
 FM - IS / I / I / ABCD / T4 Ta = +70°C
 CSA - IS Class 1 Div 1 Groups ABCD T4
 IECEx - Ex ia IIC T4 (-20°C ≤ Ta ≤ +70°C)
 
 | 
| Housing Material | 316 Stainless steel 
 | 
| Dimensions | L=132mm x Ø45mm (with connector) | 
| Filter (Sensor Protection) | Standard: HDPE Guard <10μm Optional: 316 stainless steel sintered guard <80μm
 
 | 
| Process Connection and Material 
 | 5/8"- 18 UNF Alternatives: G1/2" BSP or 3/4"- 16 UNF 316 stainless steel
 | 
| Weight | 150g 
 | 
| Interchangeability 
 | Fully interchangeable transmitter 
 | 
| Electrical Connections | Hirschmann GDS series (DIN 4350-C) 
 | 
| 
 Durum | 
| Diagnostic Conditions (Factory Programmed)
 
 | Sensor fault Under-range dew point
 Over-range dew point
 | 
| Approved Galvanic Isolators 
 | KFD2-CR-EX1.20200 KFD2-CR-EX1.30200
 KFD0-CS-EX1.50P
 KFD0-CS-EX2.50P
 
 | 
| 
 Çıkışlar | 
| Diagnostic Conditions (Factory Programmed)
 
 | 23 mA 4 mA
 20 mA
 | 
| Approved Galvanic Isolators 
 | KFD2-STC4-EX1.H MTL5041
 MTL5040
 
 |