| Performans
 | 
| Measurement Range 
 | -110 to +20°C dew point,                         0 to 1000 ppmW; -100 to +20°C dew point (for gases)          Non-standard available on request (for liquid)
 | 
| Accuracy 
 | ±1°C dew point (+20 to -60°C) ±2°C dew point (-60 to -100°C)
 
 | 
| Response Time 
 | 5 mins to T95 (dry to wet) 
 | 
| Repeatability | 0.5°C dew point 
 | 
| Elektriksel Özellikler
 | 
| Output Signal | 4-20 mA (2-wire connection, current source) User configurable over range | 
| Output | Dew point or moisture content (for gases) moisture content (for liquid) | 
| Analog Output Scaled Range 
 | Dew point: -100 to +20°C Moisture content in gas: 0-3000 ppmV
 Moisture content in liquid: 0-3000 ppmW
 Non-standard available upon request
 | 
| Supply Voltage | 12 to 28 V DC | 
| Load Resistance | Max 250 Ω @ 12 V (500 Ω @ 24 V) | 
| Current Consumption | 23 mA max, depending on output signal | 
| Saturation Constants (for moisture in liquids measurements only)
 
 | 6-point look-up table for saturation constants up to 3000 ppmW over the temperature range 0 to +50°C Saturation constants for 8 common liquids can be programmed into the Easidew PRO I.S. via the application software; Alternatively the user can program saturation constants manually | 
| CE Marked 
 | Certified 
 | 
| Çalışma Özellikleri
 | 
| Operating Temperature | -40 to +70°C 
 | 
| Operating Pressure | 45 MPa (450 barg) maximum | 
| Overpressure Rating | x2 operating pressure 90 MPa (900 barg) 
 | 
| Flow Rate 
 | 1 to 5 Nl/min mounted in standard sampling block 0 to 10 m/sec direct insertion
 
 | 
| Temperature Coefficient | Temperature compensated across operating temperature range 
 | 
| Mekanik Özellikler
 | 
| Ingress Protection 
 | IP66 in accordance with standard BS EN 60529:1992 NEMA 4 in protection accordance with standard NEMA 250-2003
 | 
| Hazardous Area Certificates | ATEX - II 1 G Ex ia IIC T4 (-20°C ≤ Ta ≤ +70°C) FM - IS / I / 1 / ABCD / T4 Ta = +70°C
 CSA - IS Class 1 Div 1 Groups ABCD T4
 IECEx - Ex ia IIC T4 (-20°C ≤ Ta ≤ +70°C)
 | 
| Housing Material 
 | 316 stainless steel | 
| Filter (sensor protection) | Standard: 316 stainless steel <80µm sintered guard Optional: <10µm HDPE guard
 | 
| Weight | 1.27kg | 
| Interchangeability 
 | Fully interchangeable transmitter | 
| Electrical Connections 
 | Screw terminal, M20 x 1.5mm | 
| Diagnostic Conditions (factory programmed)
 
 | Condition                               Output Sensor fault                            23 mA
 Under-range dew point             4 mA
 Over-range dew point               20 mA
 | 
| Approved Galvanic Isolators 
 | KFD2-CR-EX1.20200 KFD2-CR-EX1.30200
 KFD0-CS-EX1.50P
 KFD0-CS-EX2.50P
 KFD2-STC4-EX1.H
 MTL5041, MTL5040
 
 |